Analyseur d'électrolyte
Analyseur d'électrolyte
Description
Description
The electrolyte analyzer has self-test functions with hints to solve problem; 21,000 record storage; Automatic one-point and two-point calibration, plus manual calibration.
Features
- Sample type: plasma, serum, whole blood, urine.
- Self-test functions with hints to solve problem
- Wide LCD display(240*128)
- Dot matrix or thermal printer on demand
- Numeric keypad for convenient operations
- Randomly choose any combination of K,Na,Cl,Ca,pH
- High efficient cleaning procedures best for fatty samples
- Automatic calibration on alarms
- High middle and low QC values provided to adjust linearity
- RS232 port available
- Slope and intercept adjustable to ensure accuracy and linearity
- Electrode maintenance alerts
- Closed container for both calibrates and waste, which prevents the bio-organic pollution
- Auto sampler volume:39 samples
Specifications
Test range and accuracy | |||
Channel | Range | Resolution | CV% |
pH | 6.00~9.00 | 0.01 | =1.0 |
K+ | 0.50~10.00mmol/L | 0.01 | <1.0 |
Na+ | 20.0~200.0 mmol/L | 0.1 | <1.0 |
Cl– | 20.0~200.0 mmol/L | 0.1 | <1.0 |
Ca++ | 0.3~5.0mmol/L | 0.01 | <1.5 |
Li+ | 0.1~30 mmol/L | 0.01 | <1.5 |
Ambient Temperature | 5°C~40℃ |
Relative Humidity | 85% |
Sun Radiation | Avoid direct sunshine |
Electricity | 100-240v~50/60Hz |
Packing dimension | 520 * 450* 470 mm |
Net weight | 8.25Kg |
Gross weight | 10.5Kg |