Analyseur automatique d'hématologie
Analyseur automatique d'hématologie
Feature
- Throughput 60T/H.
- 14-inch touch screen.
- Tri-angle laser scatter + flow cytometry method + impedance method for RBC and PLT counting.
- 3D holographic scattergram displays the accurate 5-part differentiation of WBC.
- Large storage capacity: 100,000 results (including histogram, scattergram, patient information).
Specification
Throughput | 60Tests/hour | |||
Assay Items | 5 parts, 29parameters,3 histograms | |||
Principle | Tri-angle laser scatter, Flow cytometry method, 3D scattergram analysis, Impedance method for RBC and PLT counting, Cyanide-free method for HGB test | |||
Test Mode | CBC mode, CBC+DIFF mode Venous whole blood, Capillary whole blood and Prediluted. | |||
Parameters | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PCT, PDW, P-LCR, P-LCC, NEU%, LYM%, MON%, EOS%, BAS%, NEU#, LYM#, MON#, EOS#, BAS#
4 Research parameter: ALY%, ALY#, IG%, IG# |
|||
Performance | Item | Linearity Range | Carry Over | CV |
WBC | (0~300.0×10^9)/L | ≤0.5% | ≤2.0% | |
RBC | (0~8.00×10^12)/L | ≤0.5% | ≤1.5% | |
HGB | (0~250) g/L | ≤0.5% | ≤1.5% | |
PLT | (0~3000×10^9)/L | ≤1.0% | ≤4.0% | |
Sample Volume | CBC+DIFF mode≤20μl CBC mode≤10μl | |||
Storage | 10,000 results including histograms, scattergrams and patient information | |||
Interface | 4 USB ports, 1LAN port Bi-direction LIS, support HL7 protocol, Internal RFID reader | |||
Power Supply | AC220V±10% 50/60Hz; 110V±10% 60Hz; 400VA | |||
Package Size | 670*590*790mm(W*D*H) | |||
Gross Weight | 53kg | |||
Reagent Package Size | 320*320*290mm (W*D*H) | |||
Reagent Gross Weight | 22kg |